CHAPTER 2 DIAGNOSIS
TABLE 2-4. BALANCE CANNOT CALIBRATE USING EXTERNAL WEIGHTS
SYMPTOM PROBABLE
CAUSE
REMEDY
Balance can be turned on
but will not calibrate
manually.
Incorrect weights.
Balance is set for Legal For
Trade operation.
Balance not level.
Balance is unstable.
Verify that proper weights
are used.
Set the Setup Menu Legal
For Trade setting to Off.
(See Instruction Manual,
Sections 3.7.12 & 3.8.)
Level balance.
Perform Automatic Internal
Calibration.
->Appendix A.1.
If the balance fails to
calibrate properly, enter the
service mode (Appendix B)
and perform a Span
Calibration. If this fails,
check the InCal Weight
Assembly and the Load
Cell Assembly (Section
3.1.5 – 3.2.7). Ensure that
all parts are clean and
properly aligned.
Ohaus Corporation www.ohaus.com
2-4
Discovery Balance Service Manual
Содержание DISCOVERY BALANCE