CHAPTER 2 DIAGNOSIS
TABLE 2-2. BALANCE DOES NOT RESPOND TO FRONT PANEL CONTROLS.
SYMPTOM PROBABLE
CAUSE
REMEDY
Balance does not
respond to front panel
controls.
PC board switch or
actuating fingers on top
housing are defective.
Display PC Board is
defective.
Main PC Board is
Defective
Open the balance. (Refer to
Section 3.1.1. Using an
Ohmmeter, measure the
resistance between the pins of
the ON/Zero Off switch. It should
be open. Press the switch
through the top housing and
check that the resistance is zero.
Check each switch in a similar
manner. If continuity is not
present on a particular switch,
replace the affected switch or
actuator.
If the balance fails to turn on and
On/Zero Off switch is OK, the
Display PC Board may be
defective and should be replaced.
->Display PCB Repair &
Replacement 3.1.2
If Balance fails to turn on and
Display PCB has been replaced
then the Main PCB should be
replaced.
->Main PCB Removal &
Replacement 3.1.3
TABLE 2-3. NO DISPLAY OR PARTIAL DISPLAY
SYMPTOM PROBABLE
CAUSE
REMEDY
Display is not on or
partial characters are
displayed.
Display PC Board is
defective.
The Display PC Board is replaced
as a whole assembly. Check
procedures in Table 2-1 first and
verify that other problems do not
exist.
->Display PCB Removal &
Replacement 3.1.2.
Discovery Balance Service Manual
2-3
Ohaus Corporation www.ohaus.com
Содержание DISCOVERY BALANCE
Страница 2: ......
Страница 4: ......
Страница 10: ...TABLE OF CONTENTS Ohaus Corporation www ohaus com vi Discovery Balance Service Manual...
Страница 99: ...80104135 P N 80104135 SERVICE MANUAL DISCOVERY BALANCE...