APPENDIX A STANDARD CALIBRATION
A.1 Automatic Internal Calibration (Advanced Auto Cal)
Internal calibration can be done without using an external weight.
Note:
Be careful not to touch the scale or the table while calibration is in progress,
as it will cause the process to fail.
With the balance on, press the
Cal
button and then release it.
Automatic calibration begins. It will take a minute or so.
aUtOCAL
BUSY
When it is complete
AutoCal Done
appears.
aUtOCAL
DONE
If calibration fails,
AutoCal Abort
appears. If this
happens, level the balance and try again.
If it continues to fail, perform a Service Calibration.
If this also fails, refer to Chapter 3 of this Manual.
AUtOCAL
ABORT
When calibration is finished, the balance returns to the weighing mode.
AutoCal will automatically calibrate the balance each time there is a change in temperature
significant enough to affect accuracy, or after every 11 hours. In certain circumstances a user
may need to prevent automatic calibration, for example, while taking a series of weights.
AutoCal can be turned off from the Calibration Menu, as follows:
With the balance on, press and hold the
Cal
Menu
button until
MENU CALIBRATE
is displayed, and
CALIBRATE
is blinking. Press
Yes
.
CAL AUTOCAL
is displayed. Press
Yes
to select it.
The current
AUTOCAL
setting will then be displayed.
The AutoCal setting will be displayed as either
SET OFF
or
SET ON
.
Press
No
to toggle the display from
AUTOCAL SET ON
to
AUTOCAL SET OFF
,
then press
Yes
. Press
Exit
to exit the Calibrate Menu.
Repeat this process to re-activate AutoCal.
Automatic internal calibration is included in all Discovery models.
Ohaus Corporation www.ohaus.com
A-2
Discovery Balance Service Manual