CHAPTER 2 DIAGNOSIS
2.3 ERROR CODES
The following table lists error codes, possible causes and remedies. If the problem persists,
contact Ohaus Corporation.
TABLE 2-6. ERROR CODES
ERROR CODE
PROBABLE CAUSE
REMEDY
Error 7.0
Unstable weight reading
when defining the reference
weight
Eliminate vibration and
drafts.
Error 8.0
Weight reading below
Power On/Zero limit.
Put pan on balance.
Error 8.1
Weight reading exceeds
Power On/Zero limit.
Clear pan
Error 8.3
Weight reading exceeds
overload limit.
Clear pan.
Error 8.4
Weight reading below
underload limit.
Put pan on balance.
Error 9.0
Internal fault.
Restore EEPROM data,
using Software Service
Tool. (See Appendix C.1).
Replace Main PCB.
Error 9.5
Production calibration not
present.
Perform Service Calibration
(See Appendix B).
Error 9.8
User calibration data not
present. (Required for
LFT ON only)
Calibrate balance.
Error 53
EEPROM checksum error.
Cycle power on, off. If
balance fails to operate,
restore EEPROM data.
LOW REF WT
Average piece weight too
small. (Warning only –
count will continue.)
See Instruction Manual for
more detail.
REF WT Err
Reference weight too small.
The weight on the pan is
too small to define a valid
reference weight.
Increase sample size.
Busy
(during tare, zero, printing,
calibration, etc.)
Wait until completion.
- - - - - - - - - - -
Unstable weight reading
If balance is level, in
appropriate environment,
repair Load Cell.
Ohaus Corporation www.ohaus.com
2-6