CHAPTER 3 REPAIR PROCEDURES
3.2.5 Installing the Load Cell
4. Tighten the 3 mounting screws after aligning Load Cell. Tighten screw at point of
triangle first, then re-check alignment, and adjust as necessary. Then tighten 2
screws at base of triangle. Re-check alignment and adjust as necessary.
5. Connect the flexible cable on Load Cell, using the back of a pair of tweezers to press
cable into place. Also connect cable on Calibration Motor Assembly.
6. Place the Load Cell Cover over the Load Cell.
Figure 3-93. Replacing Load Cell Cover.
7. Note the notches for cables: top-left for Load Cell cable, and bottom-right for
Calibration Motor Assembly cable.
Figure 3-94. Load Cell with Cover in proper position,
cables resting in notches (
top-left and bottom-right
).
Ohaus Corporation www.ohaus.com
3-46
Discovery Balance Service Manual
Содержание DISCOVERY BALANCE