APPENDIX A STANDARD CALIBRATION
A.3
Calibration Test (Cal Test)
Calibration test allows a check of a known calibration mass against the last stored calibration
information in the balance.
With the balance on, press and hold
Cal Menu
until
Menu
Calibrate
is displayed with
CALIBRATE
blinking, then release the button. Press
Yes
to enter calibration.
CAL AUTOCAL
is displayed. Press
No
.
CAL
SPAN
is displayed. Press
No
.
CAL CAL TEST
is displayed. Press
Yes
and follow the
screen instructions.
CAL TST_CLEAR PAN appears for a split second.
Be sure the Pan is clear.
100_ADD WEIGHT
appears.
Place the specified weight on the Pan.
100
ADD WEIGHT
After a short period, the display indicates the difference
in weight from the last calibration, and then will display
the calibration weight on the Pan.
After the test, remove the calibration weight from the Pan.
To leave the menu, press
Exit.
Ohaus Corporation www.ohaus.com