CHAPTER 3 REPAIR PROCEDURES
3.2 Load Cell Removal and Disassembly
The Load Cell may need to be removed and repaired because of balance instability, or because
the balance does not calibrate or repeat, or because the balance is physically broken or
displays an error code. Possible causes include:
•
Parts
being
misaligned
•
Bent or broken Flexures
•
Contamination or debris in the Magnet area, or in any moving part
NOTE:
The Load Cell mechanism is extremely delicate, and must be handled very
gently in a dust-free environment after it is removed from its protective housing.
Prior to opening the Load Cell be sure to have the Service Toolset, Part No. 80030340,
which includes a fixture that must be installed to protect the Flexures during both
disassembly and assembly.
NOTE:
Always use an anti-static kit when opening the balance and handling its
electronic parts.
3.2.1 Load Cell Removal
1. Disassemble the Housing and open the balance. (Refer to Section 3.1.1.)
2. Remove the 2 screws from the Display Cable Shield Plate and lift it aside.
Display Cable Shield Plate screws
Figure 3-21. Removing Display Cable Shield Plate.
3. Carefully move the Display Cable to the side. It does not need to be disconnected at
this time.
Discovery Balance Service Manual
3-15