CHAPTER 3 REPAIR PROCEDURES
3.1.3 Main Printed Circuit Board (PCB) Removal & Replacement
8. Install the replacement PCB, carefully aligning the 3 screw holes and the RS232
screw holes.
9. Install the 2 screws that secure the RS232 connector.
10. Install the 3 screws that secure the PCB.
11. Insert the 2 small cables through the small rectangular slot on the side of the PCB
mounting plate. Connect the 3 cables back to the PCB. Be sure all 3 cables are
secured properly.
12. Re-assemble balance. (Follow Sections 3.1.2 and 3.1.3 in reverse order.)
NOTE:
Be careful when replacing the PCB Cover to avoid bending the teeth that secure the
Cover to the Mounting Plate. The Cover should rest on the inside of the rear
Mounting Plate, as shown in Figure 3-12 below.
Figure 3-12. The PCB Cover has alternating teeth to secure it
to the Mounting Plate.
13. Use the Discovery Service Tool Software to electronically install the new main PCB.
(See Appendix C.3.)
14. Perform two calibrations: first Service Calibration (see Appendix B), then Span
Calibration. (See Appendix A.2.)
15. Check functioning of the balance.