APPENDIX C SOFTWARE SERVICE TOOL INSTRUCTIONS
C.1 To Restore the EEPROM data:
1. Record the following information from the balance to be repaired:
1.1. Serial Number from the label on the back of the balance.
1.2. Model Number from the label on the back of the balance.
1.3. IDNR Number, just above the label on the back of the balance.
2. Contact Ohaus Corporation in Pine Brook NJ, and request the data file to download.
Provide the information recorded above.
3. After receiving the data file from Ohaus, continue with the following steps.
4. Connect the balance to the PC and start the Discovery Software Service Tool.
5. Click the tab labeled
Restore EEPROM
.
Figure C-2. The Software Service Tool’s Restore EEProm tab.
6. Enter the balance’s Serial Number and IDNR number.
7. Enter the path to the image file, or click the
Change
button to locate the file.
8. Click
the
Write Image File
button.
9. The software will indicate the download progress. When complete, disconnect the
power from the back of the balance, then re-connect.
10. Perform Service Calibrations (see Appendix B). Test the balance.
Discovery Balance Service Manual
C-2
Ohaus Corporation www.ohaus.com