CHAPTER 4 TESTING
4.5
Print Test
Printing data to an external computer or printer requires that the communication parameters in
the RS232 submenus be set to match external device communication parameters.
1. Remove all weight from the Pan.
2. Tare the balance, 0.0g should be displayed
3. Place a mass on the Pan.
4. Press Print button, the computer or a printer should indicate the mass value as shown
on the display.
4.6 Testing the AC Adapter
The AC Adapters are available with different input voltages. Before testing the Adapter, make
sure the Adapter rating agrees with the power source being used. All Adapters are rated with
an output of 12 V AC at 1.0 Amp. Adapters can fail by having shorted internal windings
producing low voltage output or no output at all. Measure the Voltage between Pin A and Pin B
of the Adapter (see Figure 4-6) to assure the proper voltage.
Pin A
Pin B
Note: Measure the Voltage between Pin A and Pin B
Figure 4-6. AC Adapter Cord.
Discovery Balance Service Manual
4-9
Ohaus Corporation www.ohaus.com
Содержание DISCOVERY BALANCE