OPERATING PROCEDURE
:
Optional functions
2
3
Change the internal fixation lamp position with the internal fixation lamp mark so that the
desired part of the eye is displayed.
4
Instruct the patient to close their eyes.
When the device detects that the patient’s eye is closed, the message “Please instruct the
patient to open the eyes.” appears.
5
Instruct the patient to open their eyes.
When the device detects that the patient’s eye is open, image capture is automatically exe-
cuted with flash.
6
Perform Step 7 (Page 62) to 8 of “2.3.1 Image Capture with auto shot”.
7
To capture images of the other eye, pull the joystick toward the operator, then move the
image capturing unit horizontally toward the other eye to display the anterior eye obser-
vation screen, then capture images in the same procedure as the above.
8
To perform image capture for the next patient, press the next patient button
to dis-
play the anterior eye observation screen, then perform the procedure from Step 1.
9
When image capture is complete, proceed to “2.4 Image Capture Shutdown Procedure”
(Page 67).
Содержание AFC-330
Страница 14: ...XII...
Страница 60: ...44 BEFORE USE Labels and Symbols Common See Before Use Page II...
Страница 68: ...52 BEFORE USE Device and Software Setup...
Страница 174: ...158 MAINTENANCE Consumables and Maintenance Parts List When the lamp is replaced with a spare restock the spare...
Страница 180: ...164 SPECIFICATIONS AND ACCESSORIES Configuration...
Страница 184: ...168 EMC ELECTROMAGNETIC COMPATIBILITY...