![Nidek Medical AFC-330 Скачать руководство пользователя страница 79](http://html1.mh-extra.com/html/nidek-medical/afc-330/afc-330_operators-manual_1688053079.webp)
63
OPERATING PROCEDURE
:
Image Capture (Normal Photography Mode [Auto shot, No blink])
2
If auto tracking stops before the alignment is complete
<If the auto tracking in the up, down, left, and right directions stops before the anterior eye is displayed in
the center of the screen>
The vertical movement limit indicator and the right and left movement limit indicator appear.
Operate the chinrest up/down buttons and joystick in accordance with the indicators.
<If auto tracking in the forward and backward directions stops before the alignment is complete>
The forward and backward movement limit indicator appears. Operate the joystick in accor-
dance with the indicators.
37
$EFG()JK
37
$EFG()JK
Position of the patient’s eye is too high relative to
that of the image capturing unit. Lower the chinrest.
Position of the patient’s eye is too far to the right rel-
ative to that of the image capturing unit. Move the
image capturing unit by tilting the joystick.
: Lower the chinrest.
: Raise the chinrest.
: Tilt the joystick a little to the right.
: Tilt the joystick a little to the left.
37
$EFG()JK
Forward and backward
movement limit indicator
: Tilt the joystick a little backward.
: Tilt the joystick a little forward.
Содержание AFC-330
Страница 14: ...XII...
Страница 60: ...44 BEFORE USE Labels and Symbols Common See Before Use Page II...
Страница 68: ...52 BEFORE USE Device and Software Setup...
Страница 174: ...158 MAINTENANCE Consumables and Maintenance Parts List When the lamp is replaced with a spare restock the spare...
Страница 180: ...164 SPECIFICATIONS AND ACCESSORIES Configuration...
Страница 184: ...168 EMC ELECTROMAGNETIC COMPATIBILITY...