150
MAINTENANCE
:
Error Messages and Remedies
4.2
Error Messages and Remedies
If any message shown in the list below appears in the screen, remedy the problem following the
instruction in “Cause and remedy”.
When contacting NIDEK or your authorized distributor, inform of the serial number of the device, mes-
sage number, and symptom for proper service.
Message
Cause and remedy
ERROR 001: EEPROM data error
• Abnormality in data in the backup memory (EEPROM)
Data may have become lost due to extraneous noise such as static
electricity, or the circuit board or the backup memory (EEPROM) on
the circuit board may have failed.
Turn off the device power. (See “2.4 Image Capture Shutdown
Procedure” (page 67)).
Turn on the device power again. (See “2.2 Image Capture
Preparation” (page 55)).
• If the same message appears when the device power is turned on
again, turn off the device power and contact NIDEK or your authorized
distributor.
ERROR 002: Cover open
• Abnormality was detected in the internal structure of the device.
Turn off the device power. (See “2.4 Image Capture Shutdown
Procedure” (page 67).)
Turn on the device power again. (See “2.2 Image Capture
Preparation” (page 55).)
• If the same message appears when the device power is turned on
again, turn off the device power and contact NIDEK or your authorized
distributor.
ERROR 003: The base temp. is too high
ERROR 004: Base temperature sensor error
ERROR 005: The halogen temp. is too high
ERROR 006: Halogen temp. sensor error
• The temperature inside the device may have risen suddenly.
• Turn off the device power and contact NIDEK or your authorized
distributor.
ERROR 011: Insufficient charge
ERROR 012: Charging error(4S)
ERROR 013: Charging error(1S)
ERROR 014: Flash sequence error
• The xenon flash lamp may fail.
• Turn off the device power and contact NIDEK or your authorized
distributor.
ERROR 021: QR mirror error
ERROR 023: No trigger signal detected
ERROR 024: Insufficient flashing
• Abnormality was detected in the internal structure of the device.
Turn off the device power. (See “2.4 Image Capture Shutdown
Procedure” (page 67).)
Turn on the device power again. (See “2.2 Image Capture
Preparation” (page 55).)
• If the same message appears when the device power is turned on
again, turn off the device power and contact NIDEK or your authorized
distributor.
ERROR 031: Comm error: Init. failed
• Abnormality was detected in the internal structure of the device.
• Turn off the device power and contact NIDEK or your authorized
distributor.
Содержание AFC-330
Страница 14: ...XII...
Страница 60: ...44 BEFORE USE Labels and Symbols Common See Before Use Page II...
Страница 68: ...52 BEFORE USE Device and Software Setup...
Страница 174: ...158 MAINTENANCE Consumables and Maintenance Parts List When the lamp is replaced with a spare restock the spare...
Страница 180: ...164 SPECIFICATIONS AND ACCESSORIES Configuration...
Страница 184: ...168 EMC ELECTROMAGNETIC COMPATIBILITY...