84
OPERATING PROCEDURE
:
Optional functions
2.6.6
Image Capture interval
The device is designed not to use intense light. However, although there are individual differences, the
patient]'s pupil contracts to some extent after non-mydriatic image capture. When image capture is con-
tinued on the same eye or the other eye, an interval may be needed for proper image capture.
The image capture interval display shows the approximate lapsed time after execution of image capture
by changing the color of the minimum pupil diameter mark from red to green gradually starting in the 12
o’clock direction.
This interval display functions as an hourglass.
The interval time can be set in the range from 1 to 10 minutes in 1 minute increments. The interval dis-
play can be set not to be displayed. The default setting is 1 minute.
For details of setting, see “
Photography options” (Page 113).
Example: Execution of image capture
↓
Alignment in the anterior eye observation screen for the next image capture
(The image capture interval display appears during this time.)
The speed of the interval display changes depending on the time setting.
Example: 1 minute—The red minimum pupil mark turns green in a minute.
The minimum pupil diameter
mark turns from red to green
as time passes.
Содержание AFC-330
Страница 14: ...XII...
Страница 60: ...44 BEFORE USE Labels and Symbols Common See Before Use Page II...
Страница 68: ...52 BEFORE USE Device and Software Setup...
Страница 174: ...158 MAINTENANCE Consumables and Maintenance Parts List When the lamp is replaced with a spare restock the spare...
Страница 180: ...164 SPECIFICATIONS AND ACCESSORIES Configuration...
Страница 184: ...168 EMC ELECTROMAGNETIC COMPATIBILITY...