165
6
6.
EMC
(ELECTROMAGNETIC COMPATIBILITY)
6.1
EMC (Electromagnetic Compatibility)
The device complies with the International Electrotechnical Commission standards (IEC 60601-1-2:
2007) for electromagnetic compatibility.
O Specified cable
O Essential performance
Fundus image capture function
WARNING
• Use the specified accessories, optional accessories, and cables.
Otherwise increased emissions or decreased immunity of the device may result.
• Do not use the device near, on, or under other electronic equipment.
CAUTION
• The device needs special precautions regarding EMC.
The device needs to be installed and used in accordance with the EMC information provided in this
manual.
• Portable and mobile RF communications equipment can affect the device.
Part name
Connector Shielded
Cable Shielded
Ferrite Core
Length (m)
Power cable
-
No
No
2.5
LAN cable
Yes
Yes
Yes
5.0
external fixation lamp
No
Yes
No
0.35
USB extension cable
Yes
Yes
No
0.10