153
MAINTENANCE
:
Error Messages and Remedies
4
The printer has exceeded the maximum number
of allowed simultaneous print jobs.
Please wait until other print jobs are finished.
• Although printing of multiple sheets was interrupted due to a problem
in the printer (such as shortage of printer paper or ink, or paper
jamming), there are more than five unprinted images. Only a
maximum of five images can be in the printing queue. After remedying
the problem of the printer, five images in the printing queue are
printed, but printing of the 6th and the subsequent images in the
printing queue is cancelled.
• It is recommended that the number of sheets to be printed at the same
time be limited to five.
Unable to print.
Please check the printer connection and be sure
it is connected properly.
• Printing failed even though captured images were saved with the
“Auto Print” setting set to ON.
• Check the printer power supply, printer paper, printer ink, and
connection with the device.
• For details of the Auto print setting, see “
Other settings” (page 121) in “2.9.3 External settings and Other
settings” (page 121).
Please select two or more fixation lamps.
• Setting of the fixation lamps in Panorama photography mode was 0 or
1.
In Panorama photography mode, two to nine fixation lamps need to
be selected. (See “
Pan. fixation lamps” (page 116).)
Patient ID cannot be set to hide.
• Patient ID was not selected for items to be displayed in the patient list.
Patient ID needs to be selected.
(See “
There are some patient data (CSV) that could
not be registered.
• Patient data registration using a CSV file failed.
• Check the content of the CSV file.
• When the NIDEK’s image filing software NAVIS-EX is used, patient
data that is being selected on the NAVIS-EX software or in the
Recycle Bin of NAVIS-EX cannot be registered using the CSV file.
Patient data in the Recycle Bin of NAVIS-EX
cannot be operated.
• Operation was attempted for the patient data in the Recycle Bin of
NAVIS-EX.
• Delete the patient data from the Recycle Bin or restore the patient
data to the patient list.
Unable to save data to the USB flash drive.
Check the USB flash drive.
・
At the time of disconnection from NAVIS-EX, saving of image data
that was being captured to the USB flash drive failed.
・
Check the connection of the USB flash drive.
The file cannot be transferred because a file
with the same name as 1 ) image exists in the
destination folder.
Check the output file name or the contents of
the destination folder.
・
A file with the same file name exists in the transfer destination folder.
“1)” in the message indicates the index of the image.
・
Check the setting of the output file name or the contents of the
destination folder.
The file cannot be transferred because a file
with the same name as 1 ) image exists in the
backup folder.
Check the output file name or the contents of
the backup folder.
・
A file with the same file name exists in the backup folder.
“1)” in the message indicates the index of the image.
・
Check the setting of the output file name or the contents of the
backup folder.
Any other location than the USB flash drive
cannot be referred to.
・
Display of a place other than the USB flash drive was attempted in
the “Check USB flash drive contents” window that appears by
pressing the Check contents button in the USB Flash Drive Settings
dialog.
Message
Cause and remedy
Содержание AFC-330
Страница 14: ...XII...
Страница 60: ...44 BEFORE USE Labels and Symbols Common See Before Use Page II...
Страница 68: ...52 BEFORE USE Device and Software Setup...
Страница 174: ...158 MAINTENANCE Consumables and Maintenance Parts List When the lamp is replaced with a spare restock the spare...
Страница 180: ...164 SPECIFICATIONS AND ACCESSORIES Configuration...
Страница 184: ...168 EMC ELECTROMAGNETIC COMPATIBILITY...