BEFORE USE
:
Screen Configuration
1
When disabled
Even if the patient’s eyelid is detected or the required pupil diameter is not satisfied (minimum pupil diameter
in Small pupil photography mode) (
being displayed), image capture can be executed with the auto shot
or by pressing the release button.
When enabled
The auto anti-misshooting button is displayed in color.
If the cornea-reflected spots are not detected or the pupil edge is within the required pupil diameter mark, the
required pupil diameter mark and the minimum pupil diameter mark are displayed in pink. When the cornea-
reflected spots are detected, the eyelid or eyelashes are not detected, and the pupil satisfies the required
size, the minimum pupil diameter mark is displayed in green.
If the patient’s eyelid is detected or the required pupil diameter is not satisfied (minimum pupil diameter in
Small pupil photography mode) (
being displayed), image capture cannot be executed with the auto
shot or by pressing the release button.