OPERATING PROCEDURE
:
Photography settings menu
External fixation lamp
This button is displayed only when the external fixation lamp unit cable is connected to the external fixation lamp
connector.
In Panorama photography mode, this button is not displayed and cannot be illuminated.
<Installing the external fixation lamp>
1
Attach the external fixation lamp on the left
side (as viewed from the patient) of the
headrest as shown to the right and fasten it
by tightening the knob.
2
Connect the external fixation lamp cable to
the external fixation lamp connector.
The internal fixation lamp can be used
even when the external fixation lamp unit
is connected.
<Operation of the external fixation lamp during
image capture>
1
Instruct the patient to gaze at the external
fixation lamp with the other eye.
2
While observing the fundus observation screen, move the external fixation lamp to guide
the patient’s eye to the desired direction, fix it, then capture an image.
䎨䎻䎷䎑
External fixation
lamp
Headrest
Knob
External fixation
lamp connector
Содержание AFC-330
Страница 14: ...XII...
Страница 60: ...44 BEFORE USE Labels and Symbols Common See Before Use Page II...
Страница 68: ...52 BEFORE USE Device and Software Setup...
Страница 174: ...158 MAINTENANCE Consumables and Maintenance Parts List When the lamp is replaced with a spare restock the spare...
Страница 180: ...164 SPECIFICATIONS AND ACCESSORIES Configuration...
Страница 184: ...168 EMC ELECTROMAGNETIC COMPATIBILITY...