![Nidek Medical AFC-330 Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/nidek-medical/afc-330/afc-330_operators-manual_1688053009.webp)
VII
:
CAUTION
• Each time before connecting a USB flash drive to the device, be sure to check it for
viruses.
If the device is infected with a virus and any problem occurs, NIDEK does not assume
responsibility or compensate for damages.
• Users are responsible for managing captured image data.
NIDEK will not assume any responsibility for loss of data.
• Equipment connected to the analog or digital interfaces must be certified according
to the representative appropriate national standards (such as EN 60601-1 and IEC
60601-1). Further more, all configurations shall comply with the system standard IEC
60601-1. Anyone who connects additional equipment to the signal input part or signal
output part configures a medical system, and is therefore responsible that the system
complies with the requirements of the system standard IEC 60601-1. If in doubt,
consult the technical service department of your local representative.
• This device is classified as Group 1 according to “4.4 Optical radiation hazard with
fundus cameras” in ISO 10940: 2009 Ophthalmic instruments—Fundus cameras, and
conforms to the standard.
Содержание AFC-330
Страница 14: ...XII...
Страница 60: ...44 BEFORE USE Labels and Symbols Common See Before Use Page II...
Страница 68: ...52 BEFORE USE Device and Software Setup...
Страница 174: ...158 MAINTENANCE Consumables and Maintenance Parts List When the lamp is replaced with a spare restock the spare...
Страница 180: ...164 SPECIFICATIONS AND ACCESSORIES Configuration...
Страница 184: ...168 EMC ELECTROMAGNETIC COMPATIBILITY...