MAINTENANCE
:
Error Messages and Remedies
Message with
disappears by touching the screen.
Patient ID has not been entered.
Please enter a Patient ID.
• Image capture was attempted without selecting a patient.
• Perform image capture after selecting a patient from the patient list or
after registering a new patient.
The time has been set to default.
Please set the time.
• Clock error
The time setting may have become incorrect due to a battery
shortage after an extended period of nonuse (about a month), or the
circuit board or the time IC on the circuit board may have failed.
Correct the time setting. (See “
• If the same message appears after correcting the time setting, turn off
the device power and contact NIDEK or your authorized distributor.
Unable to connect to NAVIS-EX.
• The device cannot connect to the NAVIS-EX database (PC).
• Check the LAN setting or the connection of the communication cable.
Check also whether the parameters for communication are set
correctly. Entrust the connection to service personnel of NIDEK or
your authorized distributor.
(See “
Transfer” (page 121) and “3.2.1 Network connection (LAN)”
The server has exceeded the maximum number
of allowed connections.
• The device cannot connect to the NAVIS-EX database (PC) due to the
limit of connectable devices.
• Contact NIDEK or your authorized distributor.
The destination folder for saving NAVIS-EX data
is not found.
• The destination folder for saving NAVIS-EX data cannot be not found
in the connection check.
• Check the LAN connection setting.
(See “
Unable to access to the database.
• The device cannot connect to a database (PC) other than NAVIS-EX.
• Check the LAN setting or the connection of the communication cable.
Check also whether the parameters for communication are set
correctly. Entrust the connection to service personnel of NIDEK or
your authorized distributor.
(See “
Transfer” (page 121) and “3.2.1 Network connection (LAN)”
There are some images that could not be
retrieved. The patient data may already have
been deleted or not been able to be obtained
from the database.
• Retrieval of images temporarily saved in the device was attempted
immediately after starting up the device that is used in connection with
the NAVIS-EX database (PC). However, the retrieval failed because
the patient information was not in the device.
• Check the patient information in NAVIS-EX.
Unable to register the patient data.
• Registration of patient data failed.
• Try registering the patient data again.
Unable to transfer the images.
Please check the connection settings and be
sure they are connected properly.
• Transfer of images to an external device failed.
• Check the connection to the destination external device.
(See “
Transfer” (page 121), “3.1.1 Connection, disconnection,
and use of USB flash drive” (page 140), and “3.2.1 Network
connection (LAN)” (page 144).)
Message
Cause and remedy
Содержание AFC-330
Страница 14: ...XII...
Страница 60: ...44 BEFORE USE Labels and Symbols Common See Before Use Page II...
Страница 68: ...52 BEFORE USE Device and Software Setup...
Страница 174: ...158 MAINTENANCE Consumables and Maintenance Parts List When the lamp is replaced with a spare restock the spare...
Страница 180: ...164 SPECIFICATIONS AND ACCESSORIES Configuration...
Страница 184: ...168 EMC ELECTROMAGNETIC COMPATIBILITY...