OPERATING PROCEDURE
:
Image Capture Shutdown Procedure
2.4.2
Finishing operation in order to transport the device: Packing mode
When the device is to be transported, set the device ready to be packed (by setting the image captur-
ing unit and chinrest to the positions specified for transport).
This device provides Packing mode which automatically sets the device ready to be packed.
Two methods are available to set the device to Packing mode.
From “Control menu
→
Main unit settings
→
Main unit
→
Packing mode after shutdown”, select ON or
OFF.
●
Setting device ready to be packed each time before transport
“Packing mode after shutdown”: OFF (default setting)
●
Setting device ready to be packed automatically when turning off the device power
“Packing mode after shutdown”: ON
(For the setting of Packing mode, see “
Main unit settings” (Page 105).)
Setting device ready to be packed each time before transport
This is the usual method.
1
Check that the infrared filter lever and the compensation lens select lever are fully
inserted into the device.
2
After Step 5 of “2.4.1 Usual finishing of operation” (Page 67), press the chinrest down
button
and power switch on.
3
Keep the chinrest down button
pressed until the massage “Please wait...Do not turn
off the power.” appears in the screen.
The device starts a setup for becoming ready to be packed. Wait without turning off the
device power.
4
When the message “This device is now ready to be packed. Please turn off power.”
appears, turn off the power switch.
CAUTION
• When the device is to be transported, set the device ready to be packed using
Packing mode, and pack it in the specific packing material without locking the main
unit using the locking knob.
The device may be damaged by vibration and bumping.
• If pressing of the infrared filter lever or compensation lens select lever was not sufficient in
Step 1, the setup is interrupted and a confirmation message appears.
Pressing of infrared filter lever is not sufficient: “Please check the infrared filter.”
Pressing of compensation lens select lever is not sufficient: “Please check the compensation
lens.”
Pressing the specified lever completely into the device clears the message and restarts the setup.
• To restore the device from the condition ready to be packed, turn on the power switch as in
the usual device startup.
Содержание AFC-330
Страница 14: ...XII...
Страница 60: ...44 BEFORE USE Labels and Symbols Common See Before Use Page II...
Страница 68: ...52 BEFORE USE Device and Software Setup...
Страница 174: ...158 MAINTENANCE Consumables and Maintenance Parts List When the lamp is replaced with a spare restock the spare...
Страница 180: ...164 SPECIFICATIONS AND ACCESSORIES Configuration...
Страница 184: ...168 EMC ELECTROMAGNETIC COMPATIBILITY...