69
OPERATING PROCEDURE
:
Image Capture Shutdown Procedure
2
Setting device ready to be packed automatically when turning off the device power
This method is convenient when the device is frequently packed for mass health screening or such
reasons.
1
Confirm that the infrared filter lever and compensation lens select lever are completely
pressed into the device.
2
Attach the cap on the objective lens.
3
Clean all parts (such as the chinrest and forehead rest) that came into contact with the
patients.
4
Press the shutdown button
at the top left of the anterior eye observation screen.
The shutdown confirmation message appears in the middle of the screen.
5
Press YES in the shutdown confirmation message.
The device starts a setup for becoming ready to be packed. Wait without turning off the
device power.
6
When the message “Please turn off power” appears, turn off (
) power to the device.
• If pressing of the infrared filter lever or compensation lens select lever was not sufficient in
Step 1, the setup is interrupted and a confirmation message appears
Pressing of infrared filter lever is not sufficient: “Please check the infrared filter.”
Pressing of compensation lens select lever is not sufficient: “Please check the compensation
lens.”
Pressing the specified lever completely into the device clears the message and restarts the setup.
• To restore the device from the condition ready to be packed, turn on the power switch as in
the usual device startup.
Содержание AFC-330
Страница 14: ...XII...
Страница 60: ...44 BEFORE USE Labels and Symbols Common See Before Use Page II...
Страница 68: ...52 BEFORE USE Device and Software Setup...
Страница 174: ...158 MAINTENANCE Consumables and Maintenance Parts List When the lamp is replaced with a spare restock the spare...
Страница 180: ...164 SPECIFICATIONS AND ACCESSORIES Configuration...
Страница 184: ...168 EMC ELECTROMAGNETIC COMPATIBILITY...