Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
80
Nippon Instruments Corporation
Check STD
This is used to check the status (sensitivity) of the instrument after an unknown
sample measurement.
When the sample name is set to “Check STD”, the following screen appears.
Enter the concentration and sample volume of standard mercury solution to be
measured using “Check STD”, and click “OK”.
* To enter a standard value, click the “Direct Inlet” button in the lower left part, and it
will be possible to enter a mercury quantity directly.
5.4.2.1.
START
Click
to start a measurement.
After a measurement starts, the solenoid lock operates as safety measures and it is
impossible to open the inlet until the completion of the measurement.
During a measurement, the status display lamp blinks at intervals of 0.1 sec.
Also, the following window will pop up.
* The same operation can be done by pressing “F5” key.
5.4.2.2.
STOP
When
is clicked, the cooling starts immediately and a measurement is suspended.
After the cooling is completed, the solenoid lock is disabled and it is possible to open
the inlet.
* The same operation can be done by pressing “F5” and “Shift” keys at the same time.