Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
116
Nippon Instruments Corporation
(6)
Disconnect the PharMed tube connected to the inlet.
Loosen the three screws fixing the inlet cover and inlet, and remove the inlet cover.
(7)
Pull out the sample heating tube.
Pull out the sample heating tube and mercury collector tube to the front while turning
them slowly.
(8)
Loosen the cap nut of the Teflon joint, and remove the joint.
(9)
Remove the three screws of the inlet and remove the inlet from the sample heating
tube.
Step (6)
Remove the
PharMed tube.
Step (6)
Step (7)