Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
Nippon Instruments Corporation
107
Error Messages
No.
Type
Message
1
AD
FULL
The voltage of the AD converter of the optical system
exceeds the predetermined value.
Perform “Optical Adjust”.
2
LAMP
BLOKEN
The light quantity of the lamp (REF voltage) is the
predetermined value or less.
Turn off the power and check the wiring of the lamp and
optical system.
3
CELL
DARK
The cell is dirty. Clean the cell.
4
FLOW
ERROR
The flow rate has decreased to 1/2 of the preset value or
less.
Make sure whether the piping has been clogged or not.
5
H1 THERMOCOUPLE
CONNEECT
ERROR
The thermocouple has broken.
6
H2 THERMOCOUPLE
CONNEECT
ERROR
The thermocouple has broken.
7
H3 THERMOCOUPLE
CONNEECT
ERROR
The thermocouple has broken.
8
H4 THERMOCOUPLE
CONNEECT
ERROR
The thermocouple has broken.
9
H1
BROKEN
The temperature of the H1 does not rise to the preset
temperature.
10
H3
BROKEN
The temperature of the H3 does not rise to the preset
temperature.
11
H1 OVER HEATING
Please shut down and consult technical assistance.
12
H2 OVER HEATING
Please shut down and consult technical assistance.
13
H3 OVER HEATING
Please shut down and consult technical assistance.
14
H4 OVER HEATING
Please shut down and consult technical assistance.
15
H2 HEATING ERROR
The temperature of H2 does not rise.
16
H3 HEATING ERROR
The temperature of H3 does not rise.
17
H4 HEATING ERROR
The temperature of H4 does not rise.