Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
30
Nippon Instruments Corporation
4.4.5.1.
Standard Gas
1)
Connect the collector tube as shown below.
*
Be cautious about where the activated carbon case
is connected.
Connect the tubes such that the connector
marked ”IN” becomes the upstream side.
2)
Set the pump flow rate to 0.5L/min.
3) Set the operation time of the pump to 2 minutes.
4)
Start the pump operation.
5) While the pump is in operation, inject the standard gas from MB-1 into the septum.
6)
Create the calibration curve by changing the amount of standard gas injection for each
collector tube.
Activated
carbon case
T-joint with the septum
Collector tube
Pump
Inlet
Upstream side
Downstream side
Upstream side