Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
38
Nippon Instruments Corporation
* Select the calibration curve according to the official methods.
* Select the curve (+Blank) for the measurements of trace amount of mercury close to
the blank value. So that the errors due to the difference between the intercept of the
curve and the actual blank can be eliminated.
* When a value exceeds 100 ng in the LOW range measurement, select the cubic
formula.
* When a value exceeds 1000 ng in the HIGH range measurement, select the cubic
formula.
When the “Measurement” tab in the upper left part of the screen is clicked and then
is clicked, it is possible to create and check a sample heating method.
For sample heating methods, seven types of “STD”, “SOLID, HEAVY OIL”, “AQUEOUS
LIQUID”, “COLLECTOR TUBE/ AQUA kit”, “LIGHT OIL”, “PLASTIC”, “PURGE” and
“TEST” have been prepared beforehand.
For the selection of a heating method for each sample, refer to “4.5.7.2 Registration of
Unknown Samples” in the later part of this document.
When you create a method other than those prepared beforehand, refer to “5.4.2.5
4.5.6.1.
Registration of Calibration Curve Table
Information necessary for the creation of a calibration curve is entered, and a
measurement is made.
(1)
Click the “Measurement” tab in the upper left part of the screen, and then click the
“STD” button.
(2)
Enter information necessary for the creation of a calibration curve on the following
screen: