Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
Nippon Instruments Corporation
73
(3)
The execution date of the instrument check and the value of the GLP counter are
recorded.
The items of the GLP counter are as follows.
When a threshold is exceeded, carry out component replacement.
Item
Threshold
Measures
Combustion Tube Usage
2,000 times
Replace the sample heating tube.
Adsorbed Hg [mg]
1,500 mg
Replace the activated carbon filter.
Lamp ON Time [H]
5,000 hours
Replace the lamp.
* In the case of the lamp replacement, consult with
the agency or us.
Air Pump Usage [H]
2,000 hours Replace the air pump.