Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
Nippon Instruments Corporation
21
Warning
When touching the instrument, make sure that it has been cooled
sufficiently.
When touching the instrument, put on protectors such as heat-
resistant gloves.
Otherwise, you may get burned.
Disconnect the power cable from the instrument before
maintenance work.
Otherwise, an electric shock or a failure caused by a short circuit
may result.
If the update of the firmware failed and the instrument does not start, rewrite the firmware
according to the following procedure:
(1)
Turn off the instrument and disconnect the power cable from it.
(It is unnecessary to terminate “MA3SoloWin”.)
(2)
Remove the eight screws and remove the front cover.