Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
Nippon Instruments Corporation
121
Warning
When touching the boat in/out rod, make sure that it has been
cooled sufficiently.
When touching, put on protectors such as heat-resistant gloves.
Otherwise, you may get burned.
Before using a reagent, procure and administer the safety data
sheet (SDS).
(1)
Turn the inlet knob counterclockwise while pushing up the solenoid lock locking the
boat in/out rod.
(2)
Pull out the boat in/out rod.
(3)
Remove the screw fixing the boat in/out rod, and remove it.
(4)
Remove the two screws fixing the boat in/out rod, and replace or clean it.
(5)
To clean it, cut and remove the O-ring, soak the metal part in nitric acid of 1 w/v%
or less, and carry out ultrasonic cleaning or leave it for 30 minutes.
After the cleaning, rinse it using pure water, dry it sufficiently and then attach a new
O-ring to the groove.
(6)
Tighten the two screws to fix the boat in/out rod.
* Care about the direction of the boat in/out rod.
(7)
Fix the boat in/out rod using the screw.
Step (1), (2)
Turn the inlet knob and pull out while
pushing up the solenoid lock manually.
Step (3)