Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
Nippon Instruments Corporation
131
Warning
When touching the instrument for maintenance, make sure that it
has been cooled sufficiently.
When touching, put on protectors such as heat-resistant gloves.
Otherwise, you may get burned.
Disconnect the power cable from the instrument before
maintenance work.
Otherwise, an electric shock or a failure caused by a short circuit
may result.
If “NG” is displayed for the V3 opening check in the leak check and the valve V3 may
be clogged, replace it.
(1)
Remove the connector of the valve V3.
(2)
Loosen the screws of the plate fixing the PharMed tubes, and remove the plate
and PharMed tubes from the valve.
(3)
Replace the valve and attach the components in the procedure reverse to that for
the removal.
Step (1)
Step (2)