Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
Nippon Instruments Corporation
41
4.5.6.4.
Suspension of Measurement
Warning
The area near the inlet is hot.
Put on protectors such as heat-resistant gloves, and take care not to
get burned.
When you want to suspend a measurement, click
in the upper left of the screen.
The cooling starts immediately.
After the cooling is completed, the solenoid lock is disabled.
* The same operation can be done by pressing “F5” and “Shift” keys at the same time.
4.5.6.5.
Termination of Measurement
Warning
The area near the inlet is hot.
Put on protectors such as heat-resistant gloves, and take care not
to get burned.
When handling a sample boat or a mercury collector tube for a
measurement or the like, take care not to get burned.
When handling, put on protectors such as heat-resistant gloves,
and use boat tongs.
(1)
When a measurement is completed, the following screen appears:
Also, the finishing sound will be played at the end of measurement.
* The following message will be displayed if communication stops during
measurement.
(2)
Turn the inlet knob counterclockwise and pull it toward you to take out the boat
in/out rod.
(3)
Lift the sample boat using boat tongs and disconnect the boat from the boat in/out
rod.