Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
Nippon Instruments Corporation
123
Warning
Before using a reagent, procure and administer the safety data sheet
(SDS).
Please use the following grease for the inlet.
We cannot guarantee the instrument performance if the following grease is not used.
[SUMICO LUBRICANT CO., LTD. Sumitec 103]
[SUMICO LUBRICANT CO., LTD. Sumiplex MP No.2]
(1)
According to steps (1) to (2) in “6.6.5 Replacement and Cleaning of Boat In/Out Rod”
described above, remove the boat in/out rod.
(2)
Remove grease from the following two thread parts (yellow parts).
* If the type of grease is changed, make sure to wipe off the old grease.
Mixing different types of grease can interfere the performance.
(3)
Apply grease to the thread part of the boat in/out side (yellow part) above.
Use a toothpick or the like and take care not to apply grease to any other place.
(4)
According to steps (7) and (8) in “6.6.5 Replacement and Cleaning of Boat In/Out
Rod” described above, attach the boat in/out rod as before.
the boat in/out side
the inlet side