Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
Nippon Instruments Corporation
125
Warning
Before using a reagent, procure and administer the safety data sheet
(SDS).
Caution
When touching a glass component, take care not to suffer an injury
such as cutting a hand.
Clean the removed absorption cell and joints in the following procedure:
Finally, wash the components using pure water and
then put them in a drier to dry them.
(2)
(3)
(4)
(5)
(6)
(1)
Put the absorption cell and joints in a plastic
or glass beaker.
Pour neutral detergent and clean
the components using an
ultrasonic cleaner or by keeping
them in the detergent for 30
minutes or so.
Take out and wash the
components one by one.
Pour 1:10 dilute nitric acid and clean
the components using an ultrasonic
cleaner or by keeping them in the dilute
nitric acid for 30 minutes or so.