BenchMike Pro Instruction Handbook: Setup
4-31
The
Feature
position movement mode of the Positioning Fix-
ture can be used to measure different positions in a part. In
this mode, the Collection Fixture is subordinate to the Posi-
tioning Fixture. The Positioning Fixture will move to the loca-
tion specified in the feature, and then the complete movement
of Collection Fixture will take place.
For example, you can specify a different position in three fea-
tures and select the Collection Fixture to move either
Auto-
matically
or continuously for each of those three positions.
When
Go
is pressed, the unit will start with the first feature’s
position and start measuring as the Collection Fixture is set up. When the Collection Fixture cycle is finished, the
unit will proceed to the next feature. This will continue until the last feature or, optionally, when the Feature Ad-
vance Mode can be set to
End of measurement within limits
. When configured in this manner, the unit will stop
measuring when an out of tolerance condition is encountered.
Pressing
Clear
or
Stop
before the measurement cycle is completed will stop the fixtures at the current location
and they will move to the
Home
position, if so selected.
The
Motion
tab allows you to select the desired values for the
following.
•
Motion Speed
: Sample rotation speed
•
Enable Half-Stepping
: Permits smoother sample rota-
tion
•
Home Fixture and End of Meas Cycle
: Returns fix-
ture to starting position at end of cycle
•
Disable Motor After Movement
:
Turns off fixture’s
motor at end of cycle
Содержание BenchMike Pro 2025
Страница 4: ......
Страница 14: ......
Страница 22: ......
Страница 40: ...2 18 BenchMike Pro Instruction Handbook Installation 11 Select a Make of Raw then select Add printer ...
Страница 44: ...2 22 BenchMike Pro Instruction Handbook Installation 20 Verify the report ...
Страница 50: ...2 28 BenchMike Pro Instruction Handbook Installation 10 Choose the Raw Make then click Continue ...
Страница 51: ...BenchMike Pro Instruction Handbook Installation 2 29 11 Choose the Raw Queue en Model then click Add Printer ...
Страница 112: ......
Страница 150: ......
Страница 156: ......
Страница 160: ...BenchMike Pro Instruction Handbook Index Units 4 25 Universal Slide Fixture 3 3 Ventilation Requirements 2 1 ...