6-4
BenchMike Pro Instruction Handbook: Remote Setup
Command Segment <CMD>
Each Command is composed of several message parts, separated by periods. The number of characters varies
based on the command itself. When you are asking the BenchMike Pro to return information to you, the command
ends with a question mark.
End of Transmission <EOT>
♦
The End of Transmission is a non-printable character that indicates the end of the Command Segment, or the
Returned Data.
6.2.4
Messages from the BenchMike Pro with the Wrapper
Here is the format of a message from the BenchMike Pro to the host device, with a sample message, when the
message has been understood and data is returned:
Start of Text <SOT>
☻
The Start of Text Character is a non-printable character that precedes the Status Update Character.
Status Update Character <ST>
The Status Update Character is a single ASCII character which indicates a status change. The status conditions
are grouped into four categories: System, Scanner, Flaw, and Tolerance. The Status Update Character indicates
which of these groups have changed. If you know that the status has changed, you can then request the latched
status condition from the BenchMike Pro. The following table lists the Status Update Characters in ASCII and bi-
nary formats.
Содержание BenchMike Pro 2025
Страница 4: ......
Страница 14: ......
Страница 22: ......
Страница 40: ...2 18 BenchMike Pro Instruction Handbook Installation 11 Select a Make of Raw then select Add printer ...
Страница 44: ...2 22 BenchMike Pro Instruction Handbook Installation 20 Verify the report ...
Страница 50: ...2 28 BenchMike Pro Instruction Handbook Installation 10 Choose the Raw Make then click Continue ...
Страница 51: ...BenchMike Pro Instruction Handbook Installation 2 29 11 Choose the Raw Queue en Model then click Add Printer ...
Страница 112: ......
Страница 150: ......
Страница 156: ......
Страница 160: ...BenchMike Pro Instruction Handbook Index Units 4 25 Universal Slide Fixture 3 3 Ventilation Requirements 2 1 ...