6-10
BenchMike Pro Instruction Handbook: Remote Setup
6.3
Command Descriptions
Note that the following tables list all commands in alphabetical order for simplicity. However, some commands
must be issued prior to other commands. For example, you must select the desired feature number before you
can alter the settings of that feature. A table is also given at the end of this section which lists all commands in the
order they are most commonly used.
6.3.1
Configure Settings
With the CONFigure command, you can specify settings for the serial port, features, digital inputs and outputs,
alarms, display, buttons, measurements, fixtures, and reports.
6.3.1.1
Configuring the Alarm: CONF.ALARM
The following table lists the options for configuring the alarms.
CONF.ALARM.DUR
Request or specify the duration of the alarm
signal in seconds. Set the duration value to
zero (0) to disable the alarm. Examples:
CONF.ALARM.DUR:5
CONF.ALARM.DUR?
6.3.1.2
Configuring the Buttons: CONF.BUTTON
The following table lists the options for configuring the three large data buttons on the BenchMike Pro.
NOTE
: The CONF.BUTTON.SEL command should be used to select the desired button before any MODE com-
mands are issued.
CONF.BUTTON.MODE
Request or specify the mode of operation for
the selected data button. Options include:
GO/STOP, CLEAR_Data, CLEAR_LAST, PRINT,
and SEND_DATA. Examples:
CONF.BUTTON.MODE:CLEAR_DATA
CONF.BUTTON.MODE?
CONF.BUTTON.SEL
Select the TOP, MIDDLE, or BOTTOM data
button to the right of the touchscreen. Ex-
amples:
CONF.BUTTON.SEL:TOP
CONF.BUTTON.SEL:MIDDLE
Содержание BenchMike Pro 2025
Страница 4: ......
Страница 14: ......
Страница 22: ......
Страница 40: ...2 18 BenchMike Pro Instruction Handbook Installation 11 Select a Make of Raw then select Add printer ...
Страница 44: ...2 22 BenchMike Pro Instruction Handbook Installation 20 Verify the report ...
Страница 50: ...2 28 BenchMike Pro Instruction Handbook Installation 10 Choose the Raw Make then click Continue ...
Страница 51: ...BenchMike Pro Instruction Handbook Installation 2 29 11 Choose the Raw Queue en Model then click Add Printer ...
Страница 112: ......
Страница 150: ......
Страница 156: ......
Страница 160: ...BenchMike Pro Instruction Handbook Index Units 4 25 Universal Slide Fixture 3 3 Ventilation Requirements 2 1 ...