RMX-4000 Series Programming Manual
124
Query Example
:SPEC:VOLT?
0
The test has exceeded
the Go/NoGo voltage
limits.
136B
:SPECification:VOLTage:H/L/C
Channel-Specific
Command
Description
Sets the high(H), low(L), and center(C) Go/NoGo
voltage limit specifications. Applicable to CC and
CR mode only.
Syntax
:SPECification:VOLTage:H/L/C <NRf+>[V]
Parameter
<NRf+>[V] 1 unit = 1 V
Go/NoGo voltage limit
1
1 V
1V
1 V
Example
:SPEC:VOLT:H 2V
Sets the Go/NoGo high
voltage limit to 2 V.
Query Syntax
:SPECification:VOLTage:H/L/C? <NR2>
Return Parameter <NR2>
Go/NoGo voltage
1 unit = 1 V
Returns the limit voltage
Query Example
:SPEC:VOLT:H?
2.000
The voltage limit is
2 V.
137B
:SPECification:CURRent:H/L/C
Channel-Specific
Command
Description
Sets the high(H), low(L), and center(C) Go/NoGo
current limit specifications. Applicable to CV
mode only.
Syntax
:SPECification:CURRent::H/L/C <NRf+>[A]
Parameter
<NRf+>[A] 1 unit = 1 A
Go/NoGo current limit
1
1 A
1A
1 A
Example
:SPEC:CURR:H 1A
Sets the Go/NoGo high
current limit to 1 A.
Содержание RMX-400 Series
Страница 1: ...Programmable DC Electronic Load RMX 400x Series PROGRAMMING MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 23: ...RMX 4000 Series Programming Manual 23 Memory Subsystem 153 SEQuence Subsystem 159 GLOBal Subsystem 170...
Страница 31: ...RMX 4000 Series Programming Manual 31 0 Pass 1 Fail Example TST 0...
Страница 37: ...RMX 4000 Series Programming Manual 37 Query Example MEM this is a memo Returns the memo message...
Страница 56: ...RMX 4000 Series Programming Manual 56 Query Example UTIL FRAM 0 Frame Link is on...
Страница 75: ...RMX 4000 Series Programming Manual 75 Syntax FETCh ALLPower aard...
Страница 96: ...RMX 4000 Series Programming Manual 96 1 ON Turn on Example OCP RUN OFF Turn the test off...
Страница 134: ...RMX 4000 Series Programming Manual 134 Questionable Status NTR All bits set to 0 Syntax STATus PREset Example STAT PRE...