RMX-4000 Series Programming Manual
12
Command Forms
Commands and queries have two different forms,
long and short. The command syntax is written
with the short form of the command in capitals
and the remainder (long form) in lower case.
Short
FETCh:VOLTage?
Short
Long
You can write the commands in capitals or lower
case, as long as the short or long forms are
complete. An incomplete command will not be
recognized.
Below are examples of correctly written
commands.
LONG FETCh:VOLTage?
FETCH:VOLTAGE?
fetch:voltage?
SHORT FETC:VOLT?
fetc:volt?
Square Brackets
Commands that contain squares brackets indicate
that the contents are optional. The command
function is the same with or without the square
bracketed items, as shown below.
Example:
:LOAD[:STATe]
=
:LOAD:STATe
=
:LOAD
Command
Format
1
2
3
4
:PROGram:CHAin <NR1>LF
1: Command header
2: Single space
3: Parameter
4: Message terminator
Содержание RMX-400 Series
Страница 1: ...Programmable DC Electronic Load RMX 400x Series PROGRAMMING MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 23: ...RMX 4000 Series Programming Manual 23 Memory Subsystem 153 SEQuence Subsystem 159 GLOBal Subsystem 170...
Страница 31: ...RMX 4000 Series Programming Manual 31 0 Pass 1 Fail Example TST 0...
Страница 37: ...RMX 4000 Series Programming Manual 37 Query Example MEM this is a memo Returns the memo message...
Страница 56: ...RMX 4000 Series Programming Manual 56 Query Example UTIL FRAM 0 Frame Link is on...
Страница 75: ...RMX 4000 Series Programming Manual 75 Syntax FETCh ALLPower aard...
Страница 96: ...RMX 4000 Series Programming Manual 96 1 ON Turn on Example OCP RUN OFF Turn the test off...
Страница 134: ...RMX 4000 Series Programming Manual 134 Questionable Status NTR All bits set to 0 Syntax STATus PREset Example STAT PRE...