RMX-4000 Series Programming Manual
48
35B
:CONFigure:PROTection:UVP:LEVel
Channel-Specific
Command
Description
Sets the under voltage protection level for the
current/specific channel. You can set the level to
any applicable level or to the channel
maximum/minimum.
Syntax
:CONFigure:PROTection:UVP:LEVel
<NRf>[W]|MIN|MAX
Parameter
<NRf>
UVP Level
20
20 V
20V
20 V
MIN
Sets to the minimum level
(OFF)
MAX
Sets the voltage limit to
the maximum level
Example
:CONF:PROT:UVP:LEV MIN
Sets the UVP limit to OFF.
Query Syntax
:CONFigure:PROTection:UVP:LEVel? [MIN|MAX]
Return Parameter <NRf>
Power protection level
1 unit = 1 V
Returns the UVP level as
volts
MAX/MIN
Returns the allowable
maximum and minimum
Query Example
:CONF:PROT:UVP:LEV?
75
UVP level is at 75 V.
36B
:CONFigure:RESPonse
Channel-Specific
Command
Description
Sets or queries the response rate for the specific
channel.
Syntax
:CONFigure:RESPonse {NORMAL|0|FAST|1}
Parameter
NORMAL/0
Normal
Содержание RMX-400 Series
Страница 1: ...Programmable DC Electronic Load RMX 400x Series PROGRAMMING MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 23: ...RMX 4000 Series Programming Manual 23 Memory Subsystem 153 SEQuence Subsystem 159 GLOBal Subsystem 170...
Страница 31: ...RMX 4000 Series Programming Manual 31 0 Pass 1 Fail Example TST 0...
Страница 37: ...RMX 4000 Series Programming Manual 37 Query Example MEM this is a memo Returns the memo message...
Страница 56: ...RMX 4000 Series Programming Manual 56 Query Example UTIL FRAM 0 Frame Link is on...
Страница 75: ...RMX 4000 Series Programming Manual 75 Syntax FETCh ALLPower aard...
Страница 96: ...RMX 4000 Series Programming Manual 96 1 ON Turn on Example OCP RUN OFF Turn the test off...
Страница 134: ...RMX 4000 Series Programming Manual 134 Questionable Status NTR All bits set to 0 Syntax STATus PREset Example STAT PRE...