RMX-4000 Series Programming Manual
49
FAST/1
Fast
Example
:CONF:RESP 0
Response set to normal.
Query Syntax
: CONFigure:RESPonse?
Return Parameter <NR1>
Response
0
Normal
1
Fast
Query Example
:CONF:RESP?
1
Response is Fast.
37B
:CONFigure:RESEt
Channel-Specific
Command
Description
Recalls the original factory default settings.
Syntax
:CONFigure:RESEt
Example
:CONF:RESE
38B
:CONFigure:GROup:UNITs
Channel-Specific
Command
Description
Sets or queries the number of single channel load
modules (RMX-4005 or RMX-4006) that you can
use in the parallel mode.
Syntax
CONFigure:GROup:UNITs <NRf>|MIN|MAX
Parameter
<NRf>
Number of units
MIN
Sets to the minimum
number
MAX
Sets to the maximum
number
Example
CONF:GRO:UNIT 2
Sets the parallel mode to 2
units.
Query Syntax
CONFigure:GROup:UNITs? [MIN|MAX]
Return Parameter <NR1>
Returns the number of
units
Содержание RMX-400 Series
Страница 1: ...Programmable DC Electronic Load RMX 400x Series PROGRAMMING MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 23: ...RMX 4000 Series Programming Manual 23 Memory Subsystem 153 SEQuence Subsystem 159 GLOBal Subsystem 170...
Страница 31: ...RMX 4000 Series Programming Manual 31 0 Pass 1 Fail Example TST 0...
Страница 37: ...RMX 4000 Series Programming Manual 37 Query Example MEM this is a memo Returns the memo message...
Страница 56: ...RMX 4000 Series Programming Manual 56 Query Example UTIL FRAM 0 Frame Link is on...
Страница 75: ...RMX 4000 Series Programming Manual 75 Syntax FETCh ALLPower aard...
Страница 96: ...RMX 4000 Series Programming Manual 96 1 ON Turn on Example OCP RUN OFF Turn the test off...
Страница 134: ...RMX 4000 Series Programming Manual 134 Questionable Status NTR All bits set to 0 Syntax STATus PREset Example STAT PRE...