RMX-4000 Series Programming Manual
137
10V
10 V
MIN
Sets the voltage to the
minimum value for the
channel
MAX
Sets the voltage to the
maximum value for the
channel
Example
:VOLT:AVAL 10V
Sets A Value to 10 V for
the specific channel
:VOLT:BVAL MAX
Sets B Value to the
maximum allowed
voltage for the specific
channel. (Single channel
only.)
Query Syntax
:VOLTage:AVALue/BVALue? [MAX|MIN]
Return Parameter <NR2> 1 unit = 1 V
Voltage
10
Returns the voltage of the
specific channel
MAX/MIN
Returns the allowable
maximum and minimum
Query Example
:VOLT:AVAL?
5
:VOLT:AVAL? MAX
81.6000
A Value is set to 5 V.
Returns the maximum
settable voltage.
:VOLTage:CURRent
Channel-Specific
Command
Description
Sets the current limit in CV mode.
Syntax
:VOLTage:CURRent <NRf+>[ A ]
Parameter
<NRf+>[ A ]
Current limit
1
1 A
1A
1 A
Содержание RMX-400 Series
Страница 1: ...Programmable DC Electronic Load RMX 400x Series PROGRAMMING MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 23: ...RMX 4000 Series Programming Manual 23 Memory Subsystem 153 SEQuence Subsystem 159 GLOBal Subsystem 170...
Страница 31: ...RMX 4000 Series Programming Manual 31 0 Pass 1 Fail Example TST 0...
Страница 37: ...RMX 4000 Series Programming Manual 37 Query Example MEM this is a memo Returns the memo message...
Страница 56: ...RMX 4000 Series Programming Manual 56 Query Example UTIL FRAM 0 Frame Link is on...
Страница 75: ...RMX 4000 Series Programming Manual 75 Syntax FETCh ALLPower aard...
Страница 96: ...RMX 4000 Series Programming Manual 96 1 ON Turn on Example OCP RUN OFF Turn the test off...
Страница 134: ...RMX 4000 Series Programming Manual 134 Questionable Status NTR All bits set to 0 Syntax STATus PREset Example STAT PRE...