L1/L2 MIN
Sets L1 or L2 to the
minimum level for the
specific channel
L1/L2 MAX
Sets L1 or L2 to the
maximum level for the
specific channel
Example
:RES:DYN:HIGH:L1 10
In high range CR
dynamic mode, Set L1
(level 1) to 10
Ω
.
Query Syntax
:RESistance:DYNamic:HIGH:L1/L2? [MIN|MAX]
Return Parameter <NR2> [MAX|MIN]
Resistance
1 unit= 1
Ω
Returns the resistance of
Level 1/ 2 (L1/L2)
Query Example
:RES:DYN:HIGH:L2? MAX
15000.0
Returns the maximum
resistance allowed for
the channel. (RMX-4003)
128B
:RESistance:DYNamic:HIGH:RISE/FALL
Channel-Specific
Command
Description
Sets the high range rising/falling slew rate for CR
dynamic mode for the specific channel.
Syntax
:RESistance:DYNamic:HIGH:RISE/FALL <NRf+>[A/µs]
Parameter
<NRf+>[A/µs]
Slew rate
RISE/FALL 0.8A/µs
Sets the rising/falling slew
rate to 0.8 A/µs
RISE/FALL 1
Sets the rising/falling slew
rate to 1 A/µs
RISE/FALL MIN
Sets to the slowest
rising/falling slew rate
RISE/FALL MAX
Sets to the fastest
rising/falling slew rate
Example
:RES:DYN:HIGH:RISE 1.1
Sets the rising slew rate
to 1.1 A/µs.
Содержание RMX-400 Series
Страница 1: ...Programmable DC Electronic Load RMX 400x Series PROGRAMMING MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 23: ...RMX 4000 Series Programming Manual 23 Memory Subsystem 153 SEQuence Subsystem 159 GLOBal Subsystem 170...
Страница 31: ...RMX 4000 Series Programming Manual 31 0 Pass 1 Fail Example TST 0...
Страница 37: ...RMX 4000 Series Programming Manual 37 Query Example MEM this is a memo Returns the memo message...
Страница 56: ...RMX 4000 Series Programming Manual 56 Query Example UTIL FRAM 0 Frame Link is on...
Страница 75: ...RMX 4000 Series Programming Manual 75 Syntax FETCh ALLPower aard...
Страница 96: ...RMX 4000 Series Programming Manual 96 1 ON Turn on Example OCP RUN OFF Turn the test off...
Страница 134: ...RMX 4000 Series Programming Manual 134 Questionable Status NTR All bits set to 0 Syntax STATus PREset Example STAT PRE...