RMX-4000 Series Programming Manual
111
Example
:RES:STAT:LOW:RISE 0.1
Sets the rising slew rate
to 0.1 A/µs.
Query Syntax
:RESistance:STATic:LOW:RISE/FALL? [MIN|MAX]
Return Parameter <NR2> [MAX|MIN]
Slew rate
1 unit =1 A/µs
Returns the slew rate
MAX, MIN
Returns the allowable
maximum and minimum
Query Example
:RES:STAT:LOW:RISE?
MAX
0.8000
For low range CR mode,
the maximum value for
the rising slew rate is
0.8 A/µs for the specific
channel.
122B
:RESistance:STATic:HIGH:AVALue/BVALue
Channel-Specific
Command
Description
Sets the high range A/B Value for constant
resistance static mode.
Syntax
:RESistance:STATic:HIGH:AVALue/BVALue
<NRf+>[OHM]
Parameter
NRf+[OHM]
AVALue 10
Sets A Value to 10
Ω
. (high
range only)
BVALue 20OHM
Sets B Value to 20
Ω
. (high
range only)
AVALue MIN
Sets A Value to the
minimum level for the
specific channel
AVALue MAX
Sets A Value to the
maximum level for the
specific channel
Example
:RES:STAT:HIGH:BVAL 10
Sets high range CR static
mode B Value to
10
Ω
.
Содержание RMX-400 Series
Страница 1: ...Programmable DC Electronic Load RMX 400x Series PROGRAMMING MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 23: ...RMX 4000 Series Programming Manual 23 Memory Subsystem 153 SEQuence Subsystem 159 GLOBal Subsystem 170...
Страница 31: ...RMX 4000 Series Programming Manual 31 0 Pass 1 Fail Example TST 0...
Страница 37: ...RMX 4000 Series Programming Manual 37 Query Example MEM this is a memo Returns the memo message...
Страница 56: ...RMX 4000 Series Programming Manual 56 Query Example UTIL FRAM 0 Frame Link is on...
Страница 75: ...RMX 4000 Series Programming Manual 75 Syntax FETCh ALLPower aard...
Страница 96: ...RMX 4000 Series Programming Manual 96 1 ON Turn on Example OCP RUN OFF Turn the test off...
Страница 134: ...RMX 4000 Series Programming Manual 134 Questionable Status NTR All bits set to 0 Syntax STATus PREset Example STAT PRE...