NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 133 -
o
Enter the displayed scan plane parameters from M*Vision (see Fig 8.8.14) into
the NeuroBlate Thermal Monitoring Sequences protocol’s geometry parameters
in the MRI. This will be in the
Offc/Ang
tab of the sequence exam card as shown
below. The Base plane or
slice orientation
is modified in the geometry tab as
noted.
Entering orientation values on Philips also requires the following steps in this
order to ensure final entry is correct:
•
Enter the required slice orientation (coronal/ sagittal/ transverse)
•
Enter zero for all angles (AP, RL, FH) to reset entry
•
Enter the prescribed orientation angles (AP, RL, FH)
Slice Orientation is
set here in
geometry tab
Dyn Scans Set in
Dyn/Ang tab
Slice position and
orientation is set in
offc/ang tab