NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 83 -
Figure 8.49: PCM Mounting Hub (left); PCM mounted to IP Arms at 45
°
(top right);
PCM to IP at 0° (bottom right)
•
Slide the U-shaped PCM’s mounting hub over the desired
mounting hub (figure 8.49)
•
Rotate the PCM to the desired, preset rotational positions (0°,
45° and 90°)
•
Tighten the PCM mounting hub’s locking thumbscrews until
contact is just felt between the thumbscrews and Mounting
Bracket (image right shows how far the screws should
protrude)
- DO NOT overtighten thumbscrews
PCM
Mounting
Hub
PCM Mounting
Hub Locking
Thumbscrew
PCM Hub
Locking
Thumbscrew
Underside of PCM Mounting
Hub showing the minimum
required protrusion for locking
thumbscrew engagement