The depth stop setting (or LDP ruler setting) on the LDP is derived from its use with the AXiiiS
device where the depth of the LDP is measured from the top most point of the beam direction
fiducial marker (BDM) within the AXiiiS central ball joint. This marker (shown in green in Figure
8.2.92) is visible in an MR image. When a skull bone anchor is in use, this offset must be
subtracted from L to determine the depth stop.
Figure 8.2.92: Source of Depth Stop Setting Offset (82mm)
The following table defines the LDP size needed to achieve the desired depth stop setting
computed for the various configurations:
LDP Size
Depth Stop, Min
Depth Stop, Max
000
51mm
71mm
00
72mm
92mm
1
93mm
113mm
2
114mm
134mm
3
135mm
155mm
4
156mm
176mm
5
177mm
197mm
When a RPD is intended for use with a skull anchoring device, the depth stop calculation is
different due to the longer offset the RPD adds to the bone anchor when attached. Figure
8.19c below illustrates this issue and notes the alternate computation of depth stop (or LDP
ruler setting) needed to achieve maximum target depth (PDP). The assumption here is that the
distance, L, is measured from the skull anchoring device before the RPD is attached and NOT
measured from the RPD itself which is not available in the OR during this IGS based
measurement.
PDP x
This profile on APD follower which
accepts a probe connection is identical
on all skull anchoring devices.
Содержание Neuroblate
Страница 1: ...Instructions for Use ...
Страница 23: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 23 4 2 EQUIPMENT LABELS Dornier Laser ...
Страница 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Страница 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Страница 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Страница 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...