NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 137 -
8.8.3.2.
Initiate Thermal Monitoring for all Other MRI Types
•
Start acquiring the NeuroBlate Thermal Monitoring Sequence
WARNING:
Treatment slice orientation and position must match the treatment planes
shown in the views. The MR pulse sequence parameters other than orientation and
position must
not
be changed and must adhere to the requirements for
NeuroBlate
treatment.
WARNING:
Do NOT power ON the Interface Platform during MR imaging. Image quality may
be compromised.
Note:
For any given trajectory, upon successful receipt of the very first measurement from
the MRI, the M*Vision software “locks” in the acquisition position and orientation and forces
all subsequent acquisitions to be aligned precisely to this first accepted acquisition. All
subsequent scans must have the same orientation; the position can only be shifted in equal
multiples of the slice thickness and gap. No in-plane panning or shifting is allowed after the
lock. Any incoming MR data will be rejected with a “position mismatch” error issued to the
user if precise alignment to the first accepted scan is not followed.
Содержание Neuroblate
Страница 1: ...Instructions for Use ...
Страница 23: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 23 4 2 EQUIPMENT LABELS Dornier Laser ...
Страница 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Страница 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Страница 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Страница 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...