NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 135 -
8.8.3.
Initiate Thermal Monitoring in M*Vision
•
Select the Start Acquisition button under Monitoring Status
on the M*Vision the right menu bar
This triggers acceptance of this header scan by a DICOM push
from the Philips MRI
8.8.3.1.
For Philips MRI Only (for all other MRI types skip to 8.8.3.2)
•
Once this “Header Scan” is complete, perform a DICOM push of this scan to the RTT DICOM
node (‘RTT’ in DICOM node label) only AFTER “Start Acquisition” is pressed on M*Vision
It is important to monitor the status of this transfer to ensure that it completes successfully
prior to starting the next scan (step 2). The Queue Manager can be used to monitor the
transfer completion.
For R3 or R4 MR Software:
•
Click on System menu option and then on Queue Manager…
For R5 MR Software:
•
Click on System menu option and then on Manage Job Queue…