NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 41 -
Philips MRI – FFE
Scan type
Scan Mode
Technique
Contrast enhancement
TR
TE
Flip Angle
NSA
Water Fat shift (3T) / BW (Hz)
Water Fat shift (1.5T) / BW (Hz)
FOV RL
FOV AP
FOV FH
Acq Matrix
Slice Thickness
Slice Gap
Stacks
Stacks type
Slice Scan Order
Slices
Reconstruction Matrix
Recon Voxel size
Slices per measurement
Number of measurements
Dyn Scan times
Fast Next Scan
Images
Fat and Water Suppression
Phase oversampling
Dynamic Stabilization (on 3T)
Coil Selection
Dual coil
CLEAR
Shim
Geometry Correction
Imaging
M2D
FFE
T1
27msec
19.1msec
30 deg.
1
4.346 pixels / 99.9Hz
2.175 pixels / 99.9Hz
256mm
192mm
16mm
128x196
5mm
0.25mm
1
Parallel
Interleaved
3
128
2 mm
3
512 (maximum >10000)
8.2 sec
No
1: M 2: P 3: no
None
0
Enabled
SENSE-Flex-L
No
No
Volume
Default (2D)
Note:
With Philips Flex L coils plugged in, use the default Coil selection on Philips R5 systems
(Ingenia) normally set with the “Smart Select” feature. This will select both Flex L coils and may
also set the Posterior and Q-Body coil as shown below:
Summary Panel:
00:08.2
27/19.1
Содержание Neuroblate
Страница 1: ...Instructions for Use ...
Страница 23: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 23 4 2 EQUIPMENT LABELS Dornier Laser ...
Страница 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Страница 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Страница 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Страница 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...